product name 8.4" Color TFT Fetal monitor
Features:
Single or Twins Ultrasound transducer
Data Graph and Trend Table Review of both mother and fetal
Built-in thermal printer
Built-in Network capability for connect with central monitoring software
Programmable alarms
All the colors and looks or the parameters can be customsized
Multi-face views, select focus on mother or fetal
Storage of patient information and data
Performance Specifications:
Display:8.4" Color TFT
Resolution:640x480
Display Mode:Standard View,Fetal View,Mother View
Indicator:Power indicator light, Alarm sound
Interface:Socket for connecting fetal's FHR, TOCO, and Fetal Movement sensorSocket for connecting mother's ECG, NIBP and SPO2 sensor.Net Socket(RJ45) for communication cable with Central Monitoring Station Sofrware.
Power Supply:AC 100 ~ 240V, 50/60Hz, Powe < 60VA
Trend Graph:Resolution from 1s, 5s, 10s. Maximum time 96 hours.
Trend Table:Resolution from 1s, 5s, 10s. Review up to 1000 items.
Alarm:Ajustable High and Low limits. Three level audible and vistual alarm.
User Configuration:All the color of parameters and waves can be set according to user's willing.
Technical specifications:
Safety
Meet the requirment of IEC60601 series
Type of Protection:Class II with internal electirc power supply
Degree of Protection:BF
Dimention and Weight
Dimention:92(W)x82(H)x22(D)mm
Weight:3.6kg(with battery)
Operation Environment:
Temperature:0℃~ +40℃
Humidity:15% ~ 95%
Storage Environment:
Temperature:-20℃~ +60℃
Humidity:10% ~ 95%
Patient Range:Pregnant Adult
FHR:
Ultrasound Frequency:2MHz
Range:50 ~ 210 bpm
Resolution:1bmp
Accuracy:±2bmp
TOCO:
Range:0 ~ 100%
Resolution:1%
Non-linear Error:<=±8%
Zero control:Manual
FETAL MOVEMENT:
Manual fetal movement mark
SPO2:
Measurement Range:0 ~ 100%
Resolution:1%
Accuracy:±2%(70%~100%, Adult/Pediatric, non-motion)
±3%(70%~100%, Neonate, non-motion)0% ~ 69% unspecified
Alarm Range:0% ~ 100%
Refreshing Rate:1s
Pulse Rate:
Measurement Range:25 ~ 250 bpm
Resolution:1bpm
Accuracy:±3 bpm (non-motion)
Alarm Range:25 ~ 250 bmp
Refreshing Rate:1s
ECG:
Lead mode:3-lead or 5-lead
Lead selection:I, II, III, AVR, AVL, AVF, V
HR range:15 ~ 300 bpm
ECG Waveform:2 channals
Accuracy:±2bpm or ±2%, which is greater
S-T segment detection range: -2.0mv ~ +2.0mv
Arrhythmia analysis: 13 kinds of arrhythmia
Alarm Range:15 ~ 300 bpm
NIBP:
Method:Oscillometry
Mode:Manual/Auto/Contiuous
Measurement Range:25 ~ 260mmHg
Atuo Measure Interval: 5, 10, 15, 30, 45, 60, 90 minutes
Resolution:1mmHg
Overpressure Protection: 300mmHg
Alarm range:25 ~260mmHg